SECTION 4—FRONT RIGGINGS

Adjusting Adjustable Angle Flip-Up Footplates

WARNING

When determining the angle of the footplates, make sure the rear of the footplates DO NOT interfere with the movement of the front casters. Otherwise, injury or damage may occur.

NOTE: For this procedure, refer to FIGURE 4.7.

Depth Adjustment

1. Remove the two (2) flat screws and locknuts that secure footplate to the half clamp. NOTE: Observe the angle of the footplate for reinstallation.

2. Move the footplate to one (1) of four (4) mounting positions.

NOTE: If desired depth is still not obtained, rotate the half clamp on the footplate hinge 180°.

3. Retighten the two (2) flat screws and locknuts.

NOTE: The setting for positioning the footplate on the half clamp may vary for each footplate. 4. Repeat STEPS 1-3 for opposite footplate, if desired.

Angle Adjustment

1.Loosen, but do not remove the adjustment screw in the half clamp.

2.Position the footplate to the necessary angle to accommodate the user (Detail "A"of FIGURE 4.7).

3.Retighten the adjustment screw. Torque to over 90 inch-pounds, but no more than 300 inch-pounds.

4.Repeat steps 1-3 for opposite footplate, if desired.

90° Footrest Support

Detail "A" - Side View Of

 

Footplate and Footrest Support

Flat Screws

 

Footrest Support

Footplate

Footplate

Half Clamp

Footplate

Hinge

 

Adjustment Screw

Locknuts

FIGURE 4.7 Adjusting Adjustable Angle Flip-Up Footplates

Xtra™

32

Part No 1026793

Page 32
Image 32