MWD WHEELCHAIRS

PROCEDURE 12

 

 

 

Wheelchairs Equipped With Mark IV RII Joystick (Detail "B" of FIGURE 4)

1.Attach the battery charger connector to the charger port on the FRONT of the joystick.

2.Plug the charger’s AC power cord, or extension, into the grounded 120 VAC wall outlet.

3.Wait until charging is complete.

NOTE: Allow eight (8) hours for normal charging. Larger batteries (greater than 55 ampere-hours) or severely dis- charged batteries may require up to sixteen (16) hours to be properly charged and equalized.

DETAIL "A"

Battery Charger Connector (on Side of Wheelchair)

WHEELCHAIRS

WITHOUT

SHROUDS

From Battery

Charger

Battery Charger

Connector (at Rear

REPLACING BATTERIES

WARNING

Most batteries are not sold with instructions. How- ever, warnings are frequently noted on the cell caps. Read them carefully, otherwise injury or damage can occur.

NOTE: Invacare recommends that both batteries be re- placed if one (1) battery is defective.

Recommended Battery Types

WARNING

The waranty and performance specifications contained in this manual are based on the use of deep cycle gel cell or sealed lead acid bateries. Invacare strongly recommends their use as the power source for this unit.

CAUTION

Failure to use the correct battery size and/or volt- agemaycausedamagetoyourwheelchairand give you unsatisfactory performance.

CHAIR SEAT

QTY

VOLTS

*BATTERY

REMARKS

WIDTH

 

 

SIZE

 

16,18,20-inches

2

12

22NF

Deep Cycle

 

 

 

 

 

*NOTE: RANGER II requires U1 gel cell batter-

M W D

W H E E L C H A I R S

WHEELCHAIRS WITH SHROUDS

of Wheelchair)

250 SERIES

ies.

From Battery Charger

 

DETAIL "B"

 

 

Battery Charger

Three (3)

 

 

Pronged Plug

Front View of

From Battery

 

Joystick

Charger

 

Charger

Port

Joystick

Battery

Boxes

Charger Port

FIGURE 4 - CHARGING BATTERIES

1.Remove the battery boxes from the wheelchair. Refer to REMOVING/INSTALLING BATTERY BOXES in this procedure of the manual.

2.Remove existing batteries from the battery boxes. Re- fer to INSTALLING/REMOVING BATTERIES INTO/ FROM BATTERY BOXES in PROCEDURE 10 of this manual.

3.Clean the new battery terminals.

4.Install the new batteries into the battery boxes. Refer to

INSTALLING/REMOVING BATTERIES INTO/FROM BATTERYBOXESinPROCEDURE10ofthismanual.

Cleaning Battery Terminals

WARNING

DO NOT allow the liquid in the battery to come in contact with skin, clothes or other possessions. It is a form of acid and harmful or damaging burnsmay result. Should the liquid touch your skin, wash the area immediately and thoroughly with cool water. In serious cases or if eye contact is made, seek medical attention IMMEDIATELY.

NEVERsmokeorstrikeamatchnearbatteries.Ifthe caps of the battery cells are removed, NEVER look directly into them while charging batteries.

57

Page 57
Image 57
Invacare IIBASIC, IIJR, IISTANDARD, II250 SERIES manual Detail a