SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

 

 

 

Wheelchair will not drive.

Motor release levers are

Engage motor release levers.

 

disengaged.

 

 

Batteries require charging.

Charge batteries. Make sure the setting on the

 

 

charger is correct.

 

Charger plugged in.

Unplug charger from wall outlet before

 

 

operating the wheelchair.

 

Circuit breaker tripped.

Reset circuit breaker. If breaker trips again, it

 

 

may indicate need for internal repair. Contact

 

 

Dealer/Invacare.

 

 

 

Motor “chatters” or runs

Electrical malfunction.

Contact Dealer/Invacare.

irregularly.

 

 

 

 

 

Only one drive wheel turns.

Electrical malfunction.

Contact Dealer/Invacare for service.

 

One motor lock is

Engage motor lock.

 

disengaged.

 

 

 

 

Joystick erratic or does not

Electrical malfunction.

Contact Dealer/Invacare for service.

respond as desired.

Controller programmed

Reprogram controller (Refer to electronics

 

improperly.

manual supplied with wheelchair).

 

 

 

Wheelchair does not

Poor battery terminal

Have terminals cleaned.

respond to commands.

connection.

 

 

 

 

Power indicator off - even

Electrical malfunction.

Contact Dealer/Invacare for service.

after recharging.

 

 

 

 

 

About Joystick Diagnostics

The joystick information gauge and the service indicator give indications of the type of fault or error detected by the control module. When a fault is detected, the wheelchair may stop and not drive. The LEDs on the information gauge may flash in a particular pattern or the service indicator light will flash. The number or type of flashes indicates the nature of the error. If multiple errors are found, only the first error encountered by the control module will be displayed.

Information Gauge Display Diagnostics

DISPLAY

DESCRIPTION

DEFINITION

COMMENTS

Information

Gauge

Display

All LEDs are off.

Power is off.

All LEDs are on.

Power is on.

Fewer than three LEDs on

 

 

implies reduced battery

 

 

charge.

Pronto® M51™and M61™with SureStep®

26

Part No 1125085

Page 26
Image 26