SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
MPJ+, PSR+, PSF+ Joysticks or Displays
SYMPTOM | PROBABLE CAUSE | SOLUTIONS |
|
|
|
SPM L Park Brake Fault or | Motor lock levers disengaged | Engage motor lock levers. Refer to |
SPM R Park Brake Fault displays | (Error code E9 or E10). | Engaging/Disengaging Motor Release |
| Lever on page 54. | |
and wheelchair does not drive. |
| |
|
| |
|
|
|
CHARGER PLUGGED IN displays. | Battery charger connected | Unplug battery charger from the |
| (Error code E28). | wheelchair. Refer to Charging Batteries |
|
| on page 63. |
|
|
|
SPM Battery Fault displays and the | Batteries need to be charged | Charge batteries. Refer to Charging |
wheelchair does not drive. | (Error code E14). | Batteries on page 63. If batteries fail to |
| charge properly, check battery charger | |
|
| |
|
| or replace batteries. Refer to Installing/ |
|
| Removing the Batteries on page 57. |
|
|
|
JOYSTICK TIMEOUT displays and | Joystick or input device is | Turn Off power, reconnect the joystick |
the wheelchair does not drive. | disconnected (Error code 32). | of input device and turn power On. |
|
| |
|
|
|
JS REV TOO LARGE | The joystick or input device is | Replace joystick or input device. |
JS FWD TOO LARGE | sending a value outside of the |
|
reverse, forward, left or right |
| |
JS LFT TOO LARGE or |
| |
limits (Error codes E01, E02, |
| |
JS RGT TOO LARGE | E03 or E04). |
|
displays and the wheelchair does not |
|
|
drive. |
|
|
|
|
|
NEUTRAL TESTING displays. | The joystick neutral test has | Release the joystick and try to get the |
| failed (Error code E18). | joystick back into the |
|
| position. |
|
|
|
BAD JOYSTICK CAL VALUES | The joystick calibration values | Recalibrate the joystick (joystick throw |
displays and the wheelchair does not | are outside of the expected | procedure). |
range (Error code E19). |
| |
drive. |
| |
|
| |
|
|
|
SPM NOT CONNECTED | The MPJ or Display module is | Check the connections between the |
| not communicating with the | joystick or display and the controller. |
| control module (Error code | Turn the power Off and then back On. |
| E200). | Replace the controller if necessary. |
|
|
|
SPM Communications Fault | The controller has determined a | Turn the wheelchair Off and back On. |
displays and the wheelchair drives | fault during a previous |
|
process (Error code E41). |
| |
slowly. |
| |
|
| |
|
|
|
| The Proportional or Digital | This is normal behavior. |
ATTENDANT ACTIVE and | Attendant control is active and |
|
displays. | can be used to drive the chair |
|
| (Error code W05). |
|
|
|
|
Batteries draw excessive current | Battery failure. | Have batteries checked for shorted cell. |
when charging. |
| Replace if necessary. |
| Electrical malfunction. | Contact Dealer/Invacare for service. |
|
|
|
Pronto® M91™Base with SureStep® | 30 | Part No 1143153 |