4.3.1LCD Menu Operation

The LCD display is menu driven, uses the UP, DOWN, and SELECT pushbuttons for navigation.

Press UP or DOWN to scroll through menu options.

Press SELECT to select a menu option.

When changing a variable use SELECT to move between signs or digits. When all digits flash an invalid value has been entered; press SELECT to continue

POWER UP

 

Auto-Scroll Mode

LL1

Liquid Level 1

L1C1

Level 1 Current

(Press SELECT to exit Auto-Scroll Mode)

MAIN MENU

 

 

SET

Enter Setup Menu

LL1

Liquid Level 1

L1C

Level 1 Current

DE

Output %

 

SETUP MENU

 

CAL

Enter Calibration Menu

CFG

Enter Configuration Menu

END

Exit Setup Menu

LINEARIZATION MENU

 

O01

Output Point 1

O02

Output Point 2

O20

Output Point 20

LIN SAVE

Save Linearization Table

LIN LOAD

Load Linearization Table

LIN RSET

Reset Linearization Table

END

Exit Linearization Menu

CALIBRATION MENU

LRV Set Lower Range Value (at 4mA Point)

URV Set Upper Range Value (at 20mA point)

DMP Set Damping Value (.1 to 36 Sec)

DAC TRIM † Enter DAC Trim Menu

LIN MENU Enter Linearization Menu

END Exit Calibration Menu

CONFIGURATION MENU

DE Menu (if ordered)

EUN Set Engineering Units (In, ft, mm, cm, m or %)

L1O Set Level 1 Offset

KO Module Offset

KG Module Gain

THV Threshold Voltage

RC(1) Raw Counts3

BLK Blanking Value

GS Gain Settings

LCH Latching Settings

DRC Downward Restriction Count

ALD Alarm Delay

END Exit Configuration Menu

DE Menu (optional)2

CONFIGURATION MENU 2

*To access this menu see

 

 

 

 

"Setting LTP & HTP"

DE ON/OFF

Turn DE output ON

RNG

Range Setting

 

DB ON/OFF

Extended Information

LTP

Lower Trim Point

 

END

Exit DE Menu

HTP

High Trim Point

 

 

To LL1

END

Exit Configuration Menu 2

 

 

1.Press DOWN & SELECT together for 4mA trim or UP & SELECT together for 20 mA trim

2.Available on DE output models only.

MT2000-0200-1 Rev f (10-2007) DCN0160

11