8. GLOSSARY OF TERMS
EU  | Engineering Units  | 
  | Display units selection (inches, mm, feet, %, etc.)  | 
URV  | Upper Range Value  | 
  | Location on the transmitter sensor to set as the 20 mA output  | 
LRV | Lower Range Value  | 
  | Location on the transmitter sensor to set as the 4 mA output  | 
K0  | Electronic Module Offset  | 
  | Calibration data for each unique electronics module for the MT2000  | 
KG  | Electronic Module Gain  | 
  | Calibration data for each unique electronics module for the MT2000  | 
BLK | Blanking | 
  | Area at the top of the probe that you do not want monitored for process signal return. Usually increased by a value  | 
  | of 4 per inch of nozzle.  | 
  | Default factory value of 210.  | 
THV | Threshold Voltage  | 
  | Minimum value that the return pulse voltage must be to indicate to the electronics it is a real value. Typical factory  | 
  | value to 1.5.  | 
GS  | Gain | 
  | One of four different values (1,2,3,4) which tell the electronics which algorithm to use in determining the value of the  | 
  | return pulse.  | 
HTP | High Trim Point  | 
  | Reference point near the area at which the high Engineering Units value for the transmitter will be set. To be set  | 
  | prior to LRV and URV.  | 
LTP | Low Trim Point  | 
  | Reference point near the area at which the low Engineering Units value for the transmitter will be set. To be set  | 
  | prior to LRV and URV.  | 
ULD | |
  | Method of operation of the MT2000 that allows its use in situations where there is no detectable signal along the  | 
  | length of the probe. Typically used in solids with a DC above 1.3 and below 2.0.  | 
LT  | Linearization Table | 
  | Twenty step table inside the MT2000 electronics which can be used to correct the raw signal in situations where the  | 
  | output is not linear  | 
  | Stilling Well  | 
  | Metallic Sleeve that the MT2000 probe is inserted inside. Allows probe to be used in lower DC and in situations  | 
  | where there is significant turbulence.  | 
  | Coax Probe | 
  | Probe style similar to the MT2000 installed in a Probe Well but with a much smaller diameter pipe. Can be inserted  | 
  | through a ¾” NPT fitting. Typically used on very clean fluids or compressed gas products.  | 