TESTING

Note: The control solution range

 

 

is shown in two different units of

BEFORE

measure—mg/dL (used in the United

States) and mmol/L (used in Canada

 

 

and some parts of Europe).

 

 

Do not use

after expiration date.us

 

 

 

 

 

 

 

 

 

 

 

 

4 months

after firstopening. replac

 

 

 

 

 

 

immediate ly.

 

 

 

 

 

 

 

 

 

Date vial fi rst opened __________

 

 

 

 

 

 

Lot No.

 

406814A

 

 

 

 

 

 

Exp. Date

 

07 - 96

 

 

 

 

 

 

 

ONE TOU CH Normal ControlRan

 

 

 

 

 

 

86

-

122

MG/DL

 

 

 

 

 

 

4 .

8

- 6 . 8

MMOL/L

 

 

 

 

 

 

 

 

a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(Example)

 

 

 

 

 

 

 

 

 

 

 

 

 

©LIF ESCAN INC. 1993AW

 

 

 

 

 

 

Mi lpitas,

California 95035

 

 

 

 

Control solution test results are auto- matically stored in the meter memory with the time and date of the test. If you use a logbook, indicate control solution tests by writing “C” after the result.

Control solution results

If test results fall outside the range printed on the test strip vial, repeat the test. Results that fall outside the expected range may indicate:

Procedural error

Control solution that has not been shaken enough

Expired or contaminated control solution

Improper coding of the meter

Dirt on the test strip holder or on the test area beneath the test strip holder

Test strip deterioration

Meter malfunction

Extremes in testing temperatures

24