MESSAGE

PROBLEM

WHAT TO DO

MESSAGES

error2

1. The test strip moved

retest

 

during the test.

 

2.

The test strip was not

 

 

inserted correctly.

 

3.

The test strip was

 

 

removed before the test

 

 

was completed.

 

4.

There was not enough

 

 

blood on the test strip.

 

5.

The meter was used in

 

 

very bright light.

 

6.

The check strip proce-

 

 

dure was incorrect.

 

7.

The meter may not be

 

 

operating correctly.

1.Repeat the test with a new test strip.

2.Repeat the test with a new test strip.

3.Repeat the test with a new test strip.

4.Repeat the test with a new test strip.

5.Move the meter away from the light source. Repeat the test with a new test strip.

6.Repeat the check strip test.

7.Call the LifeScan Customer Care Line at 1 800 663-5521.