MESSAGE

PROBLEM

WHAT TO DO

clean

1. There is dirt, blood, or

1. Clean the test area

test

lint on the test area.

according to instructions.

area

 

 

MESSAGES

2.Your hand or an object covered the test area while turning on the meter.

2.Repeat the test. Keep test area clear.

remove

The test strip was inserted

Remove the strip. Wait for the

strip

before the word insert

word insert before inserting

 

appeared on the display.

test strip.

 

 

 

5.6 mmol/l(Example)

If the word contrl appears

Repeat the test with a new test

contrl

after a blood test, your

strip and apply a larger drop

8:01am 6-01-95

blood sample was too small,

of blood.

 

smeared, or another drop was

 

 

added after the test began.

 

 

 

 

error 1

The sample was applied

Repeat the test with a new

retest

before the words apply

test strip.

 

sample appeared.

 