Control solution test results are automatically stored with the word contrl. (If you use a logbook, indicate control solution tests

by writing “C” after the result.)

Control solution results

If test results fall outside the range printed on the test strip vial, repeat the test. Results that fall outside the expected range may indicate:

Procedural error

Control solution that has not been shaken enough

Expired or contaminated control solution

Improper coding of the meter

Dirt on test strip holder or on test area beneath the test strip holder

Test strip deterioration

Expired or contaminated test strip

Meter malfunction

Extremes in testing temperatures

CAUTION: If your control solution test result falls outside the expected range, the system may not be working properly. Do not use the system to test your blood until you get a test result that falls within the expected range. If you are unable to resolve the problem, call your authorized LifeScan representative.

20

Page 27
Image 27
Lifescan Diabetes Monitoring System manual Control solution results