NI 4070/4072 Calibration Procedure 16 ni.com
5. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_AC_VOLTS
•Range = 0.05
•Resolution = 50e–9
6. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
7. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_AC_VOLTS_DCCOUPLED
•Range = 0.05
•Resolution = 50e–9
8. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
9. Output 50 mV at 30 Hz on the calibrator.
10. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_AC_VOLTS_DCCOUPLED
•Range = 0.05
•Resolution = 50e–9
11. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
12. Refer to Table2 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table2 for the current iteration.
b. Call niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table2 for the current iteration.
c. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
d. Call niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
e. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.