© National Instruments Corporation 17 NI 4070/4072 Calibration Procedure
13. Repeat step 12 for each of the remaining iterations shown in Table2.
14. Output 500 mV at 30 Hz on the calibrator.
15. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_AC_VOLTS_DCCOUPLED
•Range = 0.5
•Resolution = 500e–9
16. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
17. Refer to Table3 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table3 for the current iteration.
b. Call niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table3 for the current iteration.
Table 2. niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output niDMM_ConfigureMeasurement Parameters
Amplitude Frequency Function Range Resolution
150mV 50Hz NIDMM_VAL_AC_VOLTS 0.05 50e–9
50mV 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
250mV 1kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
50mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
350mV 1kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
50mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
450mV 20kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
50mV 20kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
550mV 50kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
50mV 50kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
650mV 100kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
50mV 100kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
750mV 300kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
50mV 300kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9