© National Instruments Corporation 19 NI 4070/4072 Calibration Procedure
21. Refer to Table4 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table4 for the current iteration.
b. Call niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table4 for the current iteration.
c. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
d. Call niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
e. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
22. Output 50 V at 30 Hz on the calibrator.
Table 4. niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output niDMM_ConfigureMeasurement Parameters
Amplitude Frequency Function Range Resolution
15V 50Hz NIDMM_VAL_AC_VOLTS 55e–6
5V 50Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6
25V 1kHz NIDMM_VAL_AC_VOLTS 55e–6
5V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6
35V 1kHz NIDMM_VAL_AC_VOLTS 50 50e–6
5V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
45V 1kHz NIDMM_VAL_AC_VOLTS 300 300e–6
5V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
55V 20kHz NIDMM_VAL_AC_VOLTS 55e–6
5V 20kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6
65V 50kHz NIDMM_VAL_AC_VOLTS 55e–6
5V 50kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6
75V 100kHz NIDMM_VAL_AC_VOLTS 55e–6
5V 100kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6
85V 300kHz NIDMM_VAL_AC_VOLTS 55e–6
5V 300kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6