NI 4070/4072 Calibration Procedure 32 ni.com
13. Output 20 mA on the calibrator.
14. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table19.
15. Output –20 mA on the calibrator.
16. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table19.
17. Output 200 mA on the calibrator.
18. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Resolution = 200e–9
19. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table19.
20. Output –200 mA on the calibrator.
21. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table19.
22. Output 1 A on the calibrator.
23. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_DC_CURRENT
•Range = 1
•Resolution = 1e–6
24. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table19.
25. Output –1 A on the calibrator.
26. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table19.
You have completed verifying the DC current of the NI4070/4072. Select
one of the following options:
• If you want to continue verifying other modes, go to the Verifying AC
Current section.
• If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
• If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call niDMM_close to close
the session.