© National Instruments Corporation 55 NI 4070/4072 Calibration Procedure
11. Call niDMM_CalAdjustOffset with the following parameters:
•Mode = NIDMM_VAL_AC_CURRENT
•Range = 0.01
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
12. Output 200 mA on the calibrator.
13. Call niDMM_CalAdjustGain with the following parameters:
•Mode = NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = 0.2
14. Output –200 mA on the calibrator.
15. Call niDMM_CalAdjustGain with the following parameters:
•Mode = NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = –0.2
16. Output 0 A on the calibrator.
17. Call niDMM_CalAdjustOffset with the following parameters:
•Mode = NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
18. Call niDMM_CalAdjustOffset with the following parameters:
•Mode = NIDMM_VAL_AC_CURRENT
•Range = 0.1
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
19. Output 1 A on the calibrator.
20. Call niDMM_CalAdjustGain with the following parameters:
•Mode = NIDMM_VAL_DC_CURRENT
•Range = 1
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = 1
21. Output –1 A on the calibrator.