NI 4070/4072 Calibration Procedure 56 ni.com
22. Call niDMM_CalAdjustGain with the following parameters:
Mode = NIDMM_VAL_DC_CURRENT
Range = 1
Input Resistance = NIDMM_VAL_RESISTANCE_NA
Expected Value = –1
23. Output 0 A on the calibrator with the current output set to NORM.
24. Call niDMM_CalAdjustOffset with the following parameters:
Mode = NIDMM_VAL_DC_CURRENT
Range = 1
Input Resistance = NIDMM_VAL_RESISTANCE_NA
25. Call niDMM_CalAdjustOffset with the following parameters:
Mode = NIDMM_VAL_AC_CURRENT
Range = 1
Input Resistance = NIDMM_VAL_RESISTANCE_NA
26. Call niDMM_CalAdjustMisc with Type set to
NIDMM_EXTCAL_MISCCAL_SECTION.
You have completed adjusting the current modes of the NI4070/4072.
Select one of the following options:
If you are calibrating an NI 4070, refer to one of the following sections:
Verification Procedures—to verify your new calibration coefficients
before saving them to the EEPROM
Completing the Adjustment Procedures—if you do not want to verify
the adjustments you have just made
If you are calibrating an NI4 072, refer to the Adjusting Capacitance and
Inductance (NI 4072 Only) section.
Adjusting Capacitance and Inductance (NI 4072 Only)
If you are calibrating an NI4 070, skip this section and select one of the
following options:
If you want to verify the new calibration coefficients before saving
them to the EEPROM, repeat the Verification Procedures section
(except for the Initial Setup section).
If you do not want to verify the new calibration coefficients, go to the
Completing the Adjustment Procedures section.