© National Instruments Corporation 63 NI 4070/4072 Calibration Procedure
FrequencyCapacitance and InductanceNote Because the actual capacitance verification values can differ from the following
values, Table22 provides tolerances that correspond to the NI 4072 accuracy
specifications
Table 21. Frequency Limits
NI6608 Output Frequency
2-Year Limits
Lower Upper
1Hz 0.9999Hz 1.0001Hz
20kHz 19.998 kHz 20.002kHz
500kHz 499.95kHz 500.05kHz
Table 22. NI4072 Capacitance and Inductance Verification Tolerances
Verification Values Range
2-Year Tolerances
% of Reading % of Range
0pF 300pF 0±0.5
0uH 10uH 0±1
270pF 300pF ±0.15 ±0.1
1nF 1nF ±0.15 ±0.1
1nF 10nF ±0.15 ±0.1
100nF 100nF ±0.15 ±0.1
100nF 1uF ±0.18 ±0.1
10uF 10uF ±0.18 ±0.1
10uF 100 uF ±0.18 ±0.1
1,000uF 1,000uF ±0.18 ±0.1
1,000uF 10,000uF ±0.18 ±0.1