NI 4070/4072 Calibration Procedure 20 ni.com
23. Call niDMM_ConfigureMeasurement with the following
parameters:
Function = NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 50
Resolution = 50e–6
24. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
25. Refer to Table5 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table5 for the current iteration.
b. Call niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table5 for the current iteration.
c. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
d. Call niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
e. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
Table 5. niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output niDMM_ConfigureMeasurement Parameters
Amplitude Frequency Function Range Resolution
150V 50Hz NIDMM_VAL_AC_VOLTS 50 50e–6
50V 50Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
250V 1kHz NIDMM_VAL_AC_VOLTS 50 50e–6
50V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
350V 20kHz NIDMM_VAL_AC_VOLTS 50 50e–6
50V 20kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
450V 50kHz NIDMM_VAL_AC_VOLTS 50 50e–6
50V 50kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
550V 100kHz NIDMM_VAL_AC_VOLTS 50 50e–6
50V 100kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
650V 300kHz NIDMM_VAL_AC_VOLTS 50 50e–6
50V 300kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6