© National Instruments Corporation 25 NI 4070/4072 Calibration Procedure
4. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_2_WIRE_RES
•Range = 10e3
•Resolution = 0.01
5. Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms
set to NIDMM_VAL_OFFSET_COMP_OHMS_ON.
6. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table18.
7. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_2_WIRE_RES
•Range = 1e3
•Resolution = 1e–3
8. Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms
set to NIDMM_VAL_OFFSET_COMP_OHMS_ON.
9. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table18.
10. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_2_WIRE_RES
•Range = 100
•Resolution = 100e–6
11. Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms
set to NIDMM_VAL_OFFSET_COMP_OHMS_ON.
12. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table18.
13. Remove the shorting bar from the NI4070/4072.
14. Reset the calibrator.
15. Fasten the connectors on one end of the Fluke 5440 cable to the
NI4 070/4072, and fasten the connectors on the other end of the cable
to the appropriate calibrator binding posts. Figure 1 shows the correct
connections. Table1 lists the cable connections.
16. Wait two minutes for the thermal EMF to stabilize if the Fluke 5440
cable was not previously used in this configuration.
17. Output 0 Ω on the calibrator with 2-wire compensation turned on
butwith external sense turned off.
18. Call niDMM_reset to reset the NI 4070/4072 to a known state.