NI 4070/4072 Calibration Procedure 6 ni.com
Clean any oxidation from the banana plugs on the Fluke 5440 cables
before plugging them into the binding posts of the calibrator or the
banana plug connectors of the NI 4070/4072. Oxidation tarnishes the
copper banana plugs so that they appear dull rather than shiny and
leads to greater thermal EMF.
Keep the blue banana plugs on the Fluke 5440 cables connected to the
VGUARD binding post of the calibr ator at all times.
Prevent the cables from moving or vibrating by taping or strapping
them to a nonvibrating surface. Movement or vibration causes
triboelectric effects that can result in measurement errors.
Calibration Procedures
The calibration process includes the following steps:
1. Initial Setup—Set up the test equipment.
2. Verification Procedures—Verify the existing operation of the device.
This step confirms whether the device is operating within its specified
range prior to calibration. Figure4 shows the procedural flow for
verification.
3. Adjustment Procedures—Submit the device to NI for a factory
calibration to adjust the calibration constants. Figure 5 shows the
procedural flow for adjustment.
4. Reverification—Repeat the verification procedure to ensure that the
device is operating within its specifications after adjustment.
These steps are described in more detail in the following sections.
Note In some cases, the complete calibration procedure may not be required. Refer to
Appendix A: Calibration Options for more information.

Initial Setup

Note This section is necessary for pre-adjustment verifications only. If you are performing
a post-adjustment verification, skip the setup and go directly to the Verifying DC Voltage
section.
To set up the test equipment, complete the following steps:
1. Remove all connections from the four input banana plug connectors on
the NI4 070/4072.
2. Verify that the calibrator has been calibrated within the time limits
specified in the Required Test Equipment section, and that DC zeros
calibration has been performed within the last 30 days. Consult the