NI 4070/4072 Calibration Procedure 18 ni.com
c. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
d. Call niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
e. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
18. Output 5 V at 30 Hz on the calibrator.
19. Call niDMM_ConfigureMeasurement with the following
parameters:
Function = NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 5
Resolution = 5e–6
20. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table16.
Table 3. niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output niDMM_ConfigureMeasurement Parameters
Amplitude Frequency Function Range Resolution
1500mV 50Hz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500mV 50Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
2500mV 1kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
3500mV 1kHz NIDMM_VAL_AC_VOLTS 55e–6
500mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 55e–6
4500mV 20kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500mV 20kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
5500mV 50 kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500mV 50 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
6500mV 100kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500mV 100 kH z NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
7500mV 300 kH z NIDMM_VAL_AC_VOLTS 0.5 500e–9
500mV 300kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9