NI 4070/4072 Calibration Procedure 62 ni.com
DC Current AC Current100kΩ100 kΩ±86 ppm ±17ppm
10kΩ10 kΩ±83 ppm ±14ppm
1kΩ1kΩ±83ppm ±14ppm
100Ω100 Ω±90 ppm ±25ppm
Table 19. NI 4070/4072 DC Current Verification Limits
Calibrator Amplitude Range
2-Year Limits
Lower Upper
0A 20mA –1.5μA1.5 μA
0A 200mA –4μA 4μA
0A 1A –20μA20 μA
20mA 20mA 19.989mA 20.011mA
–20mA 20 mA –20.011mA –19.989mA
200mA 200mA 199.916 mA 200.084mA
–200mA 200mA –200.084mA –199.916mA
1A 1A 0.99945A 1.00055A
–1A 1A –1.00055 A –0.99945A
Table 20. NI 4070/4072 AC Current Verification Limits
Calibrator Output
Range
2-Year Limits
Amplitude Frequency Lower Upper
1mA 1kHz 10mA 0.9976mA 1.0024mA
10mA 1kHz 10mA 9.994 mA 10.006 mA
10mA 1kHz 100mA 9.976mA 10.024 mA
100mA 1kHz 100mA 99.94 mA 100.06 mA
100mA 1kHz 1A 99.7mA 100.3mA
1A 1kHz 1A 0.9988A 1.0012 A
Table 18. NI 4070/4072 2-Wire Resistance Verification Tolerances (Continued)
Calibrator
Resistance Range
2-Year Tolerance
(ppm of Range)
24-Hour Tolerance
(ppm of Range)