© National Instruments Corporation 39 NI 4070/4072 Calibration Procedure
11. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_CAPACITANCE
•Range = 300e-12
•Resolution = 50e-15
12. Set the number of averages of the NI4072 to 20 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID = NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 20
13. Call niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table22.
Note If you use capacitance verification values that differ from the values listed in
Table10, verify that each measurement falls between the tolerances listed in Table 22.
Thet olerances shown in Table22 correspond to the NI 4072 accuracy specifications.
14. Remove the 270pF verification capacitor, and plug the 1 nF
verification capacitor into the HI and LO banana plug connectors of the
NI4 072.
15. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_CAPACITANCE
•Range = 1e-9
•Resolution = 100e-15
16. Set the number of averages of the NI4072 to 20 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID = NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 20
Note If you use capacitance verification values that differ from the values listed in
Table10, verify that each measurement falls between the tolerances listed in Table 22.
Thet olerances shown in Table22 correspond to the NI 4072 accuracy specifications.
17. Call niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table22.
18. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function = NIDMM_VAL_CAPACITANCE
•Range = 10e-9
•Resolution = 1e-12