NI 4070/4072 Calibration Procedure 24 ni.com
6. Repeat step 5 for each of the remaining iterations listed in Table8.
You have completed verifying the 4-wire resistance of the NI4070/4072.
Select one of the following options:
If you want to continue verifying other modes, go to the Verifying
2-Wire Resistance section.
If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call niDMM_close to close
the session.
Verifying 2-Wire Resistance
To verify the 2-wire resistance of the NI 4070/4072, complete the
following steps:
1. Plug in the insulated banana plug shorting bar across the HI and LO
banana plug connectors on the NI 4070/4072.
2. Wait one minute for the thermal EMF to stabilize.
3. Call niDMM_reset.
Table 8. niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator
Output
niDMM_ConfigureMeasurement Parameters
OffsetCompOhmsFunction Range Resolution
110MΩNIDMM_VAL_4_WIRE_RES 10e6 10 OFF
21MΩNIDMM_VAL_4_WIRE_RES 1e6 1OFF
3100kΩNIDMM_VAL_4_WIRE_RES 100e3 0.1 OFF
410kΩNIDMM_VAL_4_WIRE_RES 10e3 0.01 ON
51kΩNIDMM_VAL_4_WIRE_RES 1e3 1e–3 ON
6100ΩNIDMM_VAL_4_WIRE_RES 100 100e–6 ON
7 0ΩNIDMM_VAL_4_WIRE_RES 10e6 10 OFF
8 0ΩNIDMM_VAL_4_WIRE_RES 1e6 1OFF
9 0ΩNIDMM_VAL_4_WIRE_RES 100e3 0.1 OFF
10 0ΩNIDMM_VAL_4_WIRE_RES 10e3 0.01 ON
11 0ΩNIDMM_VAL_4_WIRE_RES 1e3 1e–3 ON
12 0ΩNIDMM_VAL_4_WIRE_RES 100 100e–6 ON