Press the <A> key and keeping the pressure on the key, press then the <B> key. Press in sequence, first the <A> key and then the <B> key.

9 OPERATION & APPLICATIONS

9.1Power ON

ATTENTION: ALL VALUES IN THE FOLLOWING FIGURES ARE ONLY LISTED AS AN EXAMPLE.

During set-up and load memory remember that the instructions of the manual related to key operation have the following meaning:

<A> + <B> <A> , <B>

To power the instrument press the <ON> key; the indication :

... CL526...

will appear for a few seconds.

The instrument runs an autodiagnostic routine for the self-checking of critical circuits and components.

A positive check will be shown with the indication

Test OK Ver 7.001

for about one second.

The number on the right side of the display indicates the version of the memory installed on the instrument.

The instrument is ready for measurement (IN mode) with the previously selected operating mode, as indicated below:

In 1280.6°C TcK:

Any faulty conditions that may be indicated are described in par. 8.9.

9.2Battery voltage indication

To recall the battery voltage on the display press the <SHIFT> + <BATTERY> keys. The indication will be as follows:

BAT 5.2 V IIII

The horizontal bar indicates the level of charge of the battery (each bar is equivalent to 25% of the full capacity)

The “low” limit of the battery voltage, for the correct operation of the instrument, is +4.6V.

Press any key to reset the operative mode.

During normal operating modes (measure or simulation), “low battery” condition will be shown as follows:

Out 1248.3°C TcK:

The battery symbol indicates that the battery has enough energy for about 30 minutes operation. A black symbol means that battery charge is below minimum: batteries must be recharged.

9.3Operating mode set up

To select the required operating mode follow the procedures indicated below.

24

Page 24
Image 24
Omega CL526 manual Operation & Applications, Power on, Battery voltage indication, Operating mode set up