Omega Multifunction Indicator-Simulator, CL526 manual Simulation cycle, Rj compensation mode check

Models: Multifunction Indicator-Simulator CL526

1 50
Download 50 pages 8.4 Kb
Page 31
Image 31

Press <ENTER>+<SOAK> keys to memory store the new value

Press the <MODE> key to obtain one of the following indications:

Mode 1

^ ramp

 

 

single cycle program

Mode 1

^ ramp

 

 

repeated cycling program

Mode 1

1

ramp

 

 

single cycle program

Mode +

1

ramp

repeated rise ramp program.

Press the <▲> or <▼> key to select the required program type.

Press <ENTER> + <MODE> keys to memory store the new selection.

Press the <!> or <"> key to exit the set-up procedure.

9.3.10.2Simulation cycle

Press the <SELECT> key to obtain one of the menu pages.

Press the <AUTORAMP> key to obtain one of the two following indications:

Sel : Autoramp 1

Sel : Autoramp 2

Press <▲> or <▼> key to select the required program.

Press the <AUTORAMP> to memory store the selection and to return to the menu page.

Press the <SELECT> key to return to the normal operative mode.

To run the automatic simulation cycle press the <AUTORAMP> key

The display indicates the actual cycle position as shown below:

Prg 18.0mV

On the repeated cycling to stop the program press <SHIFT> + <AUTORAMP> keys.

9.3.11Rj compensation mode check

The internal/external reference junction compensation is only enabled for temperature measurement or simulation with thermocouples.

During both simulation or measuring mode to check the type of reference junction mode previously installed press the <SELECT> key to enter the type of sensor or parameter selection menu page.

Press the <IN/OUT> key to obtain the following indication:

RJ : 22.8°C int

The above reading indicates that the instrument is preset with an internal automatic reference junction compensation. The temperature indication is the value measured by the precision thin film resistance thermometer placed inside the In/Out terminal. If, instead of the code "int" the indication