10 DIGITAL INTERFACE

The CL526 portable calibrator is equipped with a digital interface. The interface circuit is essentially based on the serial communication interface subsystem (SCI) on the chip of the microprocessor. The output voltage levels are TTL at 0 to +5 V.

An optional adaptor to convert the voltage level from 0 to +5V to RS232 levels can be supplied on request. This adaptor is required to interface the CL526 with a Personal Computer.

10.1Digital interface data program mode

To enter the procedure, press the <ENTER> + <ON> keys. The display will indicate:

CAL? 65388 N=0

To enter the program mode press the <2> key;

Baud Rate 19200

The numerical value of the “baud rate” can be one of the following : 19200, 9600, 4800, 2400, 1200, 600, 300

Select, with the <▲> or <▼> key the “baud rate” used by the receiver unit and transmission lines.

Press the <2> key to memory load the baud rate.

The display will indicate:

ID-Name : 1

The number represents the address code assigned to the instrument.

Press the <▲> or <▼> key to select a number from 00 to 99.

Press the <2> key to memory load the programmed value.

The display will return to the original indication:

CAL? 65388 N=0

To exit the procedure press the <OFF> key.

10.2Digital output wiring practice

The wiring to the digital output signals is made through a mini DIN connector mounted on the lower end of the case. The pertinent connections are indicated below.

Tx

ground

Rx

ground

5 V

Female miniDIN connector (case mounted - external view)

For easy interconnections a miniDIN connector with cable (cat. EE420123) con be supplied on request. The conductors color codes can change with different supplier; please check before using.

35

Page 35
Image 35
Omega Multifunction Indicator-Simulator, CL526 Digital Interface, Digital interface data program mode, Baud Rate, ID-Name