Omega CL526 manual Installation parameter procedure, Firmware version code Serial number

Models: Multifunction Indicator-Simulator CL526

1 50
Download 50 pages 8.4 Kb
Page 34
Image 34

Press the <ENTER> key to memory load the selection

The display will indicate the scaled input/output value.

Out

105.8 mbar

9.3.13Installation parameter procedure

To enter this procedure press the <ENTER> + <ON> keys

The display will indicate as shown below (the indicated numerical value is only an example).

CAL? 65388 N=0

To exit from the procedure, in any of the following steps switch the instrument -Off-.

9.3.13.1Firmware version code - Serial number

From the above step of the procedure it is possible to view the software version code .

Press the <IN/OUT> key to obtain the following indication:

Ver2.000 # 65535

.

The reading on the display indicates that the instrument is equipped with a memory release code 2.000.

The above information is extremely useful to understand the update status of the instrument and to simplify information exchange with OMEGA engineers during repair or service operations.

The second number on the right side of the display is the Serial Number of the instrument.

Press any key to exit the procedure with the following indication:

CAL? 65388 N=0

Switch the instrument -OFF-to end the procedure.

9.3.13.2External Rj compensation

To enter the External Rj compensation set-up start with the instrument switched -Off- and press <ENTER> + <ON> keys to obtain the following indication:

CAL? 65388 N=0

Press the <AUTORAMP> key to obtain:

RJ : 0.0°C ext

Set, with <▲> and <▼> keys, the temperature of the external Rj compensation (adjustable from -50°C to +100°C).

Press the <AUTORAMP> key to memory load the new temperature value of reference junction compensation.

To end this procedure switch the instrument -Off-

34

Page 34
Image 34
Omega CL526 manual Installation parameter procedure, Firmware version code Serial number, External Rj compensation