•Press the <ENTER> key to memory load the selection
•The display will indicate the scaled input/output value.
Out | 105.8 mbar |
9.3.13Installation parameter procedure
•To enter this procedure press the <ENTER> + <ON> keys
• The display will indicate as shown below (the indicated numerical value is only an example).
CAL? 65388 N=0
•To exit from the procedure, in any of the following steps switch the instrument
9.3.13.1Firmware version code - Serial number
From the above step of the procedure it is possible to view the software version code .
•Press the <IN/OUT> key to obtain the following indication:
Ver2.000 # 65535
.
The reading on the display indicates that the instrument is equipped with a memory release code 2.000.
The above information is extremely useful to understand the update status of the instrument and to simplify information exchange with OMEGA engineers during repair or service operations.
The second number on the right side of the display is the Serial Number of the instrument.
•Press any key to exit the procedure with the following indication:
CAL? 65388 N=0
•Switch the instrument
9.3.13.2External Rj compensation
To enter the External Rj compensation
CAL? 65388 N=0
•Press the <AUTORAMP> key to obtain:
RJ : 0.0°C ext
• Set, with <▲> and <▼> keys, the temperature of the external Rj compensation (adjustable from
•Press the <AUTORAMP> key to memory load the new temperature value of reference junction compensation.
• To end this procedure switch the instrument
34