Omega Multifunction Indicator-Simulator, CL526 manual Table of ranges and accuracy

Models: Multifunction Indicator-Simulator CL526

1 50
Download 50 pages 8.4 Kb
Page 9
Image 9

1.2.1Table of ranges and accuracy

Sensor or

Total range

High accuracy

Resolution

Limit of error

parameter

 

range

 

 

Tc type J

-210 to +1200°C

-190 to +1200°C

0.1°C

± (0.02% of rdg +0.1°C)

 

-346 to +2192°F

-310 to +2192°F

0.1°F

± (0.02% of rdg +0.18°F)

Tc type K

-270 to +1370°C

-150 to 1300°C

0.1°C

± (0.02% of rdg +0.1°C)

 

-454 to +2498°F

-220 to 2372°F

0.1°F

±(0.02% of rdg +0.18°F)

Tc type T

-270 to +400°C

-150 to 400°C

0.1°C

± (0.02% of rdg +0.1°C)

 

-454 to +752°F

-202 to 752°F

0.1°F

±0.02% of rdg +0.18°F)

Tc type R

0 to +1760°C

500 to 1700°C

0.1°C

± (0.02% of rdg +0.3°C)

 

32 to +3200°F

932 to 3092°F

0.1°F

± (0.02% of rdg +0.154°F)

Tc type S

0 to +1760°C

600 to 1760°C

0.1°C

± (0.02% of rdg +0.3°C)

 

32 to +3200°F

1112 to 3200°F

0.1°F

± (0.02% of rdg +0.154°F)

Tc type B

200 to +1820°C

1000 to 1820°C

0.1°C

±(0.02% of rdg +0.4°C)

 

392 to +3308°F

1832 to 3308°F

0.1°F

± (0.02% of rdg +0.172°F)

Tc type C

0 to +2300°C

1150 to 2300°C

0.1°C

± (0.02% of rdg +0.4°C)

 

32 to +4172°F

2102 to 4172°F

0.1°F

± (0.02% of rdg +0.172°F)

Tc type G

0 to +2300°C

300 to 2000°C

0.1°C

± (0.02% of rdg +0.4°C)

 

32 to +4172°F

572 to 3632°F

1°F

± (0.02% of rdg +0.172°F)

Tc type D

0 to +2300°C

300 to 2000°C

0.1°C

± (0.02% of rdg +0.4°C)

 

32 to +4172°F

572 to 3632°F

0.1°F

± (0.02% of rdg +0..72°F)

Tc type U

-200 to +400°C

-150 to +400°C

0.1°C

±(0.02% of rdg +0.1°C)

 

-328 to +752°F

-238 to +752°F

0.1°F

± (0.02% of rdg +0.18°F)

Tc type L

-200 to +760°C

-200 to 760°C

0.1°C

± (0.02% of rdg +0.1°C)

 

-328 to +1400°F

-328 to 1400°F

0.1°F

± (0.02% of rdg +0.18°F)

Tc type N

0 to 1300°C

0 to 1300°C

0.1°C

±(0.02% of rdg +0.1°C)

 

+32 to 2372°C

32 to 2372

0.1°F

± (0.02% of rdg +0.18°F)

Tc type E

-270 to +1000°C

-200 to +1000°C

0.1°C

±(0.02% of rdg +0.1°C)

 

-454 to +1832°F

-328 to +1832°F

0.1°F

± (0.02% of rdg +0.18°F)

Tc type F

0 to 1400°C

0 to 1400°C

0.1°C

±(0.02% of rdg +0.1°C)

 

32 to 2552°F

32 to 2552°F

0.1°F

± (0.02% of rdg +0.18°F)

Pt 100 (IEC)

-200 to +850°C

-200 to +850°C

0.1°C

±(0.02% of rdg +0.1°C)

 

-328 to +1562°F

-328 to +1562°F

0.1°F

± (0.02% of rdg +0.18°F)

Pt100 (JIS)

-200 to +600°C

-200 to +600°C

0.1°C

±(0.02% of rdg +0.1°C)

 

-328 to +1562°F

-328 to +1562°F

0.1°F

± (0.02% of rdg +0.18°F)

Pt100 (US)

-200 to +850°C

-200 to +850°C

0.1°C

±(0.02% of rdg +0.1°C)

 

-328 to +1562°F

-328 to +1562°F

0.1°F

± (0.02% of rdg +0.18°F)

Ni100

-60 to +180°C

-60 to +180°C

0.1°C

±(0.02% of rdg +0.1°C)

 

 

-76 to +356°F

0.1°F

± (0.02% of rdg +0.18°F)

Ni120

0 to +150°C

0 to +150°C

0.1°C

±(0.02% of rdg +0.1°C)

 

32 to +302°F

32 to +302°F

0.1°F

± (0.02% of rdg +0.18°F)

mV

-18.000 to +22.000

-10.000 to +22.000

1 µV

± (0.01% of rdg +3µV)

mV

-10.00 to +100.00

-10.000 to +21.000

10 µV

± (0.01% of rdg +3µV)

 

 

+21.00 to +53.00

10 µV

± (0.01% of rdg +3µV)

 

 

+53.00 to +100.00

10 µV

± (0.01% of rdg +6µV)

mV

0.0 to +1000.0

0.0 to +1000.0

100 µV

± (0.01% of rdg +40µV)

V

0.000 to +10.000

0.000 to +10.000

1 mV

± (0.02% of rdg +0.4mV)

 

 

 

 

 

mA

0.000 to +21.000

0.000 to +21.000

1 µA

± (0.02% of rdg +0.5 µA)

(IN)

0.0 to 400.0

0.0 to 400.0

10 m

± (0.02% of rdg +38m)

(OUT)

0.0 to 400.0

0.0 to 400.0

10 m

± (0.03% of rdg +78m)

Note:

Accuracy shown are based on tests at 23°C ±5°C for 360 days

All Input ranges: additional error ± 1 digit

Traceability chart to WECC or SIT available on request

9