1.2Specifications

Target dimensions at different distances

Nominal target diameter at 95% energy

Temperature measuring ranges: -30 to 2000°C (-22 to 3632°F)

Thermocouple measuring ranges:

type K : -100°C to +1370°C (-148 to 2500°F) 0.1°C resolution

type S : 0°C to +1760°C (-32 to 3200°F) 0.1°C resolution

Spectral response:

OSP 500, 800 & 1000 : 8 - 14 µm

OSP 1300 : 1.6 µm

OSP 1600 & 2000 0.9 µm

Accuracy:

OSP 500, 800 & 1000: ±(1% of the reading +1°C/2°F)

OSP 1300, 1600 & 2000: ±(0.5% of the reading +1°C/2°F)

Repeatability IR:

OSP 500, P800 & P1000 : ±0.5% of the reading

OSP 1300, P1600 & P2000: ±0.25% of the reading

Temperature stability IR:

for the band exceeding +18°C to +28°C: ±0.01% of f.s./°C

Display:

High contrast custom LCD with backlight device

Display resolution:

1°C / °F / K (0.1°C/0.1°F in AVG mode up to 200°C)

Alarms:

Acoustic and visible

Measurement sampling time: <300 ms

Emissivity:

Adjustable from 0.10 to 1.00 in 0.01 steps

Analog output:

1mV/ °C/ °F/ K

Digital interface:

full bi-directional TTL level port Optional TTL to RS232 adapter

Calculated functions: Hold, average, max, min, T

Data memory:

500input data structured by tag

Calibration:

self learning technique with automatic procedure

Laser:

Wavelength: 650nm

Maximum optical power: <1mW

FDA Classification: Class II, Complies with 21CFR Chapter 1, Subchapter J

Safety classification: Class 2

Beam diameter: 3mm

Beam divergence: <0.5mrad

Laser indicator: asterisk on display

Power supply:

alkaline or rechargeable type AA battery

Battery life:

16h (with backlight off)

Battery low level of charge: symbol on the LCD display

Line operation:

100, 115, 230 Vac through the external charger

Charger transformer insulation:

2500 V

Battery recharging time:

8h at 90% (instrument switched off)

Operating environment temperature range: from -10°C to +55°C (-14 to 131°F)

Storage temperature range:

from -30 °C to +60 °C (-22 to 140°F) without battery

Case:

Injection moulded ABS+ policarbonate

Dimensions:

200x 180 x 80 mm (7.87 x 7.09 x 3.15”)

Weights:

net 0.8 kg (1.76 lb); gross 1.5 kg (3.3 lb)

8

Page 8
Image 8
Omega OSP SERIES manual Specifications, Thermocouple measuring ranges