9.9 Automatic Emissivity SettingThis function is used to make an automatic setting of emissivity using an external thermocouple connected with the
auxiliary input.
• If necessary, enable the thermocouple input (see par. 9.2) and select the thermocouple type (see par. 8.6)
• Press the <▲> or <▼> key until the thermocouple temperature measure is displayed
e
°C
Lock
• Press <ENTER> key for setting the emissivity value with reference to the temperature values read from thermocouple
and pyrometer. The emissivity will change automatically to the correct value by the OSP.
e
°C
Lock
• An error message appear if the OSP measure a temperature highest of the thermocouple. If the error happen, verify
the contact between the thermocouple and the object.
e
°C
Lock