Reading values

$00 - Pyrometer measure --------------------------------------------------------------

RX

DATA1

STATUS

 

 

DATA2

-

 

 

DATA3

VALUE (H)

 

 

DATA4

VALUE (L)

 

 

IF STATUS=0

MEASURE = SIGNED INTEGER ( VALUE ) / 10

 

IF STATUS=2

MEASURE = ERROR RJ

 

IF STATUS=10

MEASURE = OFL

 

IF STATUS=11

MEASURE = UFL

 

$01 - Emissivity----------------------------------------------------------------------------

RX

DATA1

-

 

DATA2

-

 

DATA3

-

 

DATA4

VALUE

 

EMISSIVITY = VALUE / 100

 

$03 - Max input value --------------------------------------------------------------------

RX

DATA1

-

 

DATA2

-

 

DATA3

VALUE (H)

 

DATA4

VALUE (L)

 

MAX = SIGNED INTEGER ( VALUE ) / 10

 

$04 - Min input value---------------------------------------------------------------------

RX

DATA1

-

 

DATA2

-

 

DATA3

VALUE (H)

 

DATA4

VALUE (L)

 

MIN = SIGNED INTEGER ( VALUE ) / 10

 

$05 - Average input value --------------------------------------------------------------

RX

DATA1

-

 

DATA2

-

 

DATA3

VALUE (H)

 

DATA4

VALUE (L)

 

AVERAGE = SIGNED INTEGER ( VALUE ) / 10

 

$06 - Tag ID,Eng.unit,Tc type,IPTS -------------------------------------------------

RX

DATA1

Tag ID (0=Tag0, 1=TagA, 2=TagB,.....,19=TagS)

 

DATA2

Eng.unit (1=°C - 2=K - 3=°F)

 

DATA3

Tc type (0=Tc K - 1=Tc S)

 

DATA4

IPTS (0=68 - 1=90)

 

$07 - Tc measure --------------------------------------------------------------------------

RX

DATA1

STATUS

 

 

DATA2

-

 

 

DATA3

VALUE (H)

 

 

DATA4

VALUE (L)

 

 

IF STATUS=0

MEASURE = SIGNED INTEGER ( VALUE ) / 10

 

IF STATUS=1

MEASURE = ERROR RJ

 

IF STATUS=2

MEASURE = TC OFF

 

IF STATUS=10

MEASURE = OPTC

 

$08 - HAL Setpoint------------------------------------------------------------------------

RX

DATA1

-

 

DATA2

-

 

DATA3

VALUE (H)

 

DATA4

VALUE (L)

 

HAL = SIGNED INTEGER ( VALUE )

 

IF HAL=3001 THEN HAL=OFF

 

$09 - LAL Setpoint ------------------------------------------------------------------------

RX

DATA1

-

 

DATA2

-

 

DATA3

VALUE (H)

 

DATA4

VALUE (L)

 

LAL = SIGNED INTEGER ( VALUE )

 

IF LAL=3001 THEN LAL=OFF

 

$0B - First 4 chars tag description--------------------------------------------------

RX

DATA1

'X' (ASCII CODE)

 

DATA2

'X' (ASCII CODE)

 

DATA3

'X' (ASCII CODE)

 

48

Page 48
Image 48
Omega OSP SERIES manual $00 Pyrometer measure, Reading values