6 PRE-OPERATIONAL CHECK

OSP series portable thermometers are powered either by four alkaline or by Ni-MH rechargeable batteries (optional). The external battery charger, supplied as a standard , is set for 115Vac power source.

Before using the instrument carefully verify the nominal voltage value of the charger with the available mains power line. The instrument should be used in environments where the temperature does not exceed the specified limits (from -5°C to +50°C) and where the relative humidity is lower than 95%.

In case of “low” battery conditions (voltage lower than 4.6 V) the display will show the appropriate symbol. A battery symbol means that the battery package has enough energy for about 20 minutes operation. In this condition the instrument batteries must be recharged.

WARNING

USE OSP IR PORTABLE THERMOMETERS TO MEASURE TEMPERATURE: 0.9 - 1.1 M (OSP1600, OSP2000), 1.6 M (OSP1300)

AND 8-14 M (OSP800, OSP1000 AND OSP500) SPECTRAL BANDS.

IN CASE IR THERMOMETER SHOULD MEASURE DIRECT OR REFLECTED HIGH INTENSITY RADIATIONS IN THE ABOVE SPECTRAL BAND,

THEY CAN DAMAGE PERMANENTLY THE IR SENSORS INSIDE THE THERMOMETERS.

THIS KIND OF RADIATION CAN BE PRODUCED BY EITHER THE LASER TYPE ND :YAG (λEM = 1.06 M) OR BY THE LASER TYPE CO2

(λEM = 10.6 M).