Orion 2xx manual 5.1.2Status Messages, Display Message, Action, Comment

Models: 2xx

1 36
Download 36 pages 17.03 Kb
Page 23
Image 23
5.1.2Status Messages

5.1.2Status Messages

Display Message

Action

Comment

OK

None

Normal operating mode

 

 

 

Initial

None

Shown at power-up during

 

 

self check

 

 

 

Default Cal

Factory set default calibration parameters

Consult Factory

 

are in use, level reading may be inaccurate

 

 

 

 

Lo Temp

Present temperature in electronics

Transmitter may need to be moved

 

compartment is below -40° C

to ensure temperature is within

 

 

specification

 

 

 

Hi Temp

Present temperature in electronics

Transmitter may need to be moved

 

compartment is above +80° C

to ensure temperature is within

 

 

specification

 

 

 

Float 2

No level signal detected from float 2

Make sure 2 floats are being used,

Fail

 

are not damaged, and within

 

 

measuring range

 

 

 

Float 1

No level signal detected from float 1

Make sure float is not damaged

Fail

 

and within measurement range

 

 

 

No Signal

No signal detected from any float

Make sure float is not damaged

 

 

and within measurement range

 

 

 

Snsr Brd Fail

No signal from probe, bad board or

Check electrical connections

 

connection problem

to probe

 

 

 

DfltParm

Internal non-volatile, parameters have

Consult Factory

 

been defaulted

 

 

 

 