5.3 Troubleshooting

5.3 Troubleshooting

PLP-601 / PLP-602 / PLP-601R / PLP-602R

Rotor

Display

 

Content

Checking point

 

 

 

 

 

 

 

Display lamps are

1

Battery capacity

1

Replace with fresh

 

not turned on.

 

is low.

 

batteries or charge

 

 

 

 

 

 

 

battery

 

 

 

 

2

Incorrect battery

2

Check the polarity

 

 

 

 

 

insertion.

 

of the batteries.

 

 

 

 

 

 

 

 

 

Lamp

1

turns on.

1

Instrument’s tilt

1

Turn off the power

 

 

 

 

 

exceeds auto leveling

 

supply and reset the

 

 

 

 

 

area. (± 5.7°)

 

instrument level

 

 

 

 

2

Instrument is tilt

 

(± 5.7°) and turn on

Does not rotate

 

 

 

 

largely by any cause.

 

again.

 

 

 

3

After power supply

2

Turn off the power

 

 

 

 

 

 

 

 

 

turns on, unstable

 

supply and turn on

 

 

 

 

 

state of auto

 

again.

 

 

 

 

 

self-leveling function

3

Turn off the power

 

 

 

 

 

continues

 

supply and turn on

 

 

 

 

 

approx.3 min.

 

again.

 

 

 

 

 

NOTE*

 

 

 

 

 

 

 

 

 

Lamp

1

& lamp 2

Defect of instrument

Turn off the power

 

blink mutually

itself.

supply and turn on

 

 

 

 

 

 

again. If the lamps blink

 

 

 

 

 

 

farther, instrument need

 

 

 

 

 

 

to be repaired.

 

 

 

 

 

 

 

 

*NOTE:

When the operation is continued, set the laser beam on the reference mark and perform it.

Manual mode display lamp: 2

 

 

OFF

 

 

LEVEL

 

 

Off level display lamp: 1

MANUAL

BATTERY

 

MODE

CHECK

POWER

 

 

ON

 

 

OFF

37

Page 37
Image 37