Remote Technologies RPC-320 Differential Mode, Examples u sing CON FIG AIN, Acquiring Analog Data

Models: RPC-320

1 51
Download 51 pages 34.03 Kb
Page 37
Image 37

ANALOG INPUT

following tables for differ ential inputs.

When channel = odd

Pol.

-

+

-

+

-

+

-

+

CH #

0

1

2

3

4

5

6

7

channel

1

 

 

3

5

 

 

7

When channel =

even

 

 

 

 

 

Pol.

+

-

+

-

+

-

+

-

CH #

0

1

2

3

4

5

6

7

channel

0

 

 

2

4

 

 

6

When range = 0, the input is ±2.5 v olts and a 1 = 0 to 5 volts.

Differential Mode

When d ifferential m ode is specified , inpu ts are actu ally pseudo-differential. What this means is that a ground reference is needed. For example, you cannot place a battery be tween ch annel 0 and 1 and get an ac curate reading. T he (-) input must be referenced to ground. An example of where pseudo-differential works is an output from a bridge network.

Apseudo-differential input subtracts the DC component from an input. T he IC maker recomm ends the (-) input remain stable within 1 count with respect to ground for best results. Connecting a 0. 1 uF capacitor from the (-) input to grou nd wor ks well.

When operating in differential mode, r elative + and - voltages must be connected to specific inputs. When inputs are reversed, a conversion returns a 0. When the relative voltage changes, perform a conversion on the alternate c hannel. CON FIG AIN is p erfor med on b oth channels.

Pairs of channels c an be differ ential while oth ers single ended. Thus, if channel 0 and 1 are differ ential inputs, channels 2-7 may be single ended.

Examples u sing CON FIG AIN

Below are sample syntaxes for CO NFIG AIN

Differential, 0 to + 5V input

CON FIG AIN 0,0, 1

CON FIG AIN 1,0, 1

SECTION 10

Per form a conver sion as nor mal:

A = AIN(0)

The difference between channel 0 and 1 is returned. When channel 1 is more positive than channel 0, the result is zero. The differenc e is read on channel 1 by performing:

A = AIN(1)

Single-ended, ±2.5V input

CONFIG AIN channel, 1 , 0

The result is 0 for -2.500V input, 2048 for 0. 000V, and 4095 for + 2.4988V.

Acquiring Analog Data

Analog data is accessed with the AIN function. The syntax is:

A = AIN(channel)

This function assigns the analog value of a channel to the variable ; A in th is case. The value retur ned is alway s in the 0 to 4095 range bec ause the converter is 12 bits.

Power up or reset default configures inputs to the 0-5V range, single ended.

To view the result of a conversion in the command mode, type:

print ain(0)

The result at channel 0 is returned. T he returned value will always be in the 0 to 4095 range. When using a channel in the ±2. 5V ra nge, the value r eturned is interpreted differently. Zero count is now -2.500V, 4095 is + 4.9988, and 2048 is 0.000V.

Use the following formulas to convert a returned number to a voltage:

0 - 5V A = . 001221 * AIN(channel)

±2.5V A = .001221 * ain( channel) - 2.5

The AIN function require s about 1.5 ms to convert the data. Additional time is needed to store the data. The example below takes 255 data samples and stores them

Page 10-2 RPC -320