Replacing the Air Filter

Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of a VPAP Malibu unit, the air filter needs to be replaced every

six months (or more often if your unit is in a dusty environment). To replace the air filter:

Note: For details of how to check and replace the hypoallergenic filter (if used) see the Hypoallergenic Air Filter User Instructions.

1. Remove the air filter cover at the back of the VPAP Malibu.

Air filter cover

2.Remove and discard the old air filter.

3.Insert a new filter with the blue tinted side facing out.

4.Replace the air filter cover.

WARNING

! Do not wash the air filter. The air filter is not washable or reusable.

Replacing the Antibacterial Filter

You should regularly inspect and replace the antibacterial filter, as outlined in the Antibacterial Filter User Instructions.

Note: ResMed recommends that you use the Pall BB50TE filter.

Servicing

This product (VPAP Malibu) should be inspected by an authorized ResMed Service Center 5 years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an Authorized ResMed Service Center.