268106r1.book Page 35 Thursday, November 30, 2006 2:35 PM

Technical Specifications

Performance

Operating pressure range: 3 to 25 cm H2O

Maximum single fault pressure: 40 cm H2O

Pressure measurement tolerance: ±0.5 cm H2O + 4% of the measured reading

Flow measurement tolerance: ±0.2 L/s

Dynamic pressure characteristics

VSet mode: 4 to 25 cm H2O (measured at the end of standard 6’6" air tubing);

AutoSet Max pressure: 20 cm H2O; Pressure Support: 0 to 10 cm H2O

S mode: IPAP: 3 to 25 cm H2O (measured at the end of standard 6’6" air tubing);

EPAP: 3 to 25 cm H2O (measured at the end of standard 6’6" air tubing)

CPAP mode: 4 to 20 cm H2O (measured at the end of standard 6’6" air tubing)

Dynamic flow characteristics

130 L/min at 3 to 25 cm H2O

Sound pressure level: <30 dB (tested in accordance with the requirements of ISO 17510-1:2002)

Data storage: 365 days of usage, leak, pressure, AHI and AI data

Dimensions (L x W x H): 10.6" x 9.1" x 5.6"

Weight: 5.1lb

Power supply: Input range 100–240V, 50–60Hz, 40VA (typical power consumption), < 100VA (maximum power consumption)

Housing construction: Flame retardant engineering thermoplastic

Environmental conditions

Operating temperature: 41oF to 104oF (+5oC to +40oC)

Operating humidity: 10–95% non-condensing

Storage and transport temperature: -4oF to 140oF (-20oC to +60oC)

Storage and transport humidity: 10–95% non-condensing

Electromagnetic Compatibility

Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments. For further details, see “Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity” on page 37.

Air filter: Two-layered, powder-bonded, polyester non-woven fiber

Air tubing: Flexible plastic, 1 x 6’6" (2 m) or 1 x 9’10" (3 m) length

Air outlet: 22mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory Equipment - Conical Connectors

IEC 60601-1 classification: Class II (double insulated), Type CF

Note: The manufacturer reserves the right to change these specifications without notice.

Technical Specifications

English

35

Page 40
Image 40
ResMed C16389 dimensions Technical Specifications, Performance