268106r1.book Page 6 Thursday, November 30, 2006 2:35 PM

WARNING

! This is NOT a life support device. It may stop operating with power failure or if a fault occurs in the unit.

The air flow for breathing produced by this device can be as much as 11oF (6oC) higher than the temperature of the room. Caution should be exercised if the room temperature is warmer than 90oF (32oC).

The VPAP Malibu unit should only be used with masks (and connectors)

recommended by ResMed, or by your physician or respiratory therapist. A mask should not be used unless the VPAP Malibu unit is turned on and

operating properly. The vent hole or holes associated with the mask should never be blocked.

Explanation: The VPAP Malibu unit is intended to be used with special masks (or connectors) which have vent holes to allow continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP or bilevel devices.

The VPAP Malibu unit should only be connected to the components, humidifiers, or accessories specified in this guide. Connection of other items may result in injury, or damage to the VPAP Malibu unit.

At low pressures, the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing. Some rebreathing may occur.

Explosion hazard—do not use in the vicinity of flammable anesthetics.

If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating.

Explanation of the warning: When the device is not in operation and the oxygen flow is left on, oxygen delivered into the ventilator tubing may accumulate within the device enclosure. Oxygen accumulated in the device enclosure will create a risk of fire. This warning applies to most types of flow generators.

Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open flame.

Always ensure air flow is being generated by the device before the oxygen supply is turned on.

Always turn the oxygen supply off before stopping the air flow from the device.

Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary, depending on where the oxygen is introduced, pressure settings, patient breathing pattern, mask selection, and leak rate. This applies to most types of CPAP machines.

6

Page 11
Image 11
ResMed C16389 dimensions 268106r1.book Page 6 Thursday, November 30, 2006 235 PM