Problem

Possible Cause

Solution

 

 

 

 

There is excessive leak.

Adjust position of mask and

 

 

headgear.

 

 

Plugs may be missing from

 

 

ports on mask. Replace

 

 

them.

 

 

Air tubing not connected

 

 

properly. Connect firmly at

 

 

both ends.

 

 

Air tubing kinked or

 

 

punctured. Straighten or

 

 

replace.

 

 

 

VPAP Malibu unit does

Mask Alarm has been

Disable Mask Alarm to

not stop when you

enabled; SmartStop has

enable SmartStop.

remove your mask.

automatically been

 

 

disabled.

 

 

Use of a full face mask.

SmartStop does not work

 

 

with a full face mask.

 

 

 

SmartStop is enabled

Incompatible humidifier or

Use only equipment as

but the flow generator

mask system being used.

recommended and supplied

does not stop

 

by ResMed.

automatically when

 

 

you remove your mask.

 

 

 

 

 

Display error message:

The air tubing is loose.

Check tube!!

Key if done

Check that the air tubing is connected securely to your mask and the air outlet on the front of the

VPAP Malibu.

Air tubing is blocked.

The device motor may not be running at the correct speed..

Check air tubing for blockage.

Return your unit for servicing.

To clear the error message, press any key on the VPAP Malibu keypad.

32