English

Power-off

If the sensors are strained or you have called up a menu, the scale will not power off until the ON/Off key has been pressed for 3 seconds.

The scale cannot be powered off while waiting for an input for editing in setting mode. To power-off you must exit the editing area.

OFF display

In the Service/General/OFF display setting mode you can set whether

the word "OFF" is to appear in the display when the scale is switched off (medical directive).

Zero

Zeroing corrects minor deviations from the zero point caused, for example, by soiling of the scale.

The following symbol in the symbol bar indicates zero: >0<.

Zero range calibratable and non calibratable: -1% to +3% of the weighing range. If zeroing is not possible, the error message

"Above zero range" or "Below zero range" appears for two seconds.

Zero tracking

Zero tracking automatically corrects minor deviations from the zero display.

x 10 display

Indicates the weight value with an additional decimal in 10 times higher resolution.

+

Switch to x10 display by pressing first the Print key and then simultaneously the

Zeroing key.

 

According to the Calibration Law, the x10 function on a calibratable scale must only be activated briefly for checking purposes.

For bed weighing, the x10 function is programmed to switch off again automatically 5 seconds after being activated.

While the ten times higher display is activated, X10 appears in the symbol bar. When the appliance is switched off, the function is deactivated.