4Ć34
This procedure shows the setups and lists the steps to check the Sweep
RateAccuracy.

%&" %! (  '"% &#)

  
High
Frequency
Signal
Generator
W


W 
Output


#!&# %! (  '"% &#)

+%" +Install the SDĆ24 Sampling Head into the compartment shown
in the setupdiagram (CH 1/CH 2 slot). If the unit was in standby mode
set the ON/STANDBYswitch to ON.
+%" +Initialize the mainframe as follows:
UTILITYbutton press...........................................
 %* (in major menu) touch..............................
 %* (in verification popĆup menu) touch..................
+%" +Set the horizontal record length as follows:
!#*! % $ (in major menu) touch.......................
 !#  %(in popĆup menu) touch.................
Set  !#  to 1024 points adjust upper knob.........
+%" +Identifythe trace as follows:
 #icon touch.............................................
Identify Channel () touch...................................
Select  %# $ touch....................................
'" % &#)