Corrective Maintenance
CSA 803C Service Manual 6Ć35
A5 Time Base/Controller BoardA6 Calibrator Board
Removal and replacement steps are listed below.See Figures 6Ć10 and 6Ć32
forconnector, screw, and index locations.
ăStep 1:ăPlace the instrument in the inverted position.
ăStep 2:ăToremove the A6 Calibrator Board, remove connectors J1 and
J17 from the A6 Calibrator Board. Remove the Torxhead screw securing
the board in place. Lift the board from J91 on the A5 TimeBase/ControlĆ
ler Board. Perform this process in reverse to replace the Calibrator
Board.
ăStep 3:ăRemove connectorsJ16, J29A , J30A, J32, J35, J91, Ext A, and
Presc (if Prescaler is present) from the A5 TimeBase/Controller board.
Use extreme care to prevent damage to the cable center connector that
was attached to Presc. Then remove connectors J18 and J83 from the
A5 Time Base/Controller board. Note the position of the multiĆpin conĆ
nector's index triangles to ensure that the connectors can be correctly
replaced.
NOTE
         
        
  
ăStep 4:ăRemove the six Torxhead screws holding the board in place.
ăStep 5:ăRemove the A5 Time Base/Controller board.