Performance Tests
CSA 803C Service Manual 4Ć29
ăStep 3:ăTriggerthe CSA 803C as follows:
TRIGGERbutton press.........................................
Source (in major menu) touch...............................
Internal Clock (in popĆup menu) touch.......................
ăStep 4:ăDefinethe trace as follows:
Def Traicon touch.............................................
Identify Channel (2) touch...................................
Select Enter Desc touch....................................
ăStep 5:ăSet Averaging as follows:
Acquire Desc (in major menu) touch.............................
Average N (in popĆup menu) touch..............................
Select Exit touch..............................................
ăStep 6:ăSelect the Mean Measurement as follows:
MEASUREButton press........................................
Measurements(in major menu) touch........................
Mean (in popĆup menu) touch...............................
Compare & Reference (in major menu) touch.................
Save Current Meas Valuesas References
(in popĆup menu) touch.....................................
Compare off (changes to on / in popĆup menu) touch..........
Select Exit touch..........................................
ăStep 7:ăSet the VerticalSize and Offset as follows:
Verticalsize icon touch.........................................
 the VertSize: M2 to 10 mV/div adjust upper knob........
VertOffset: M2 (in major menu) touch........................
Keyin 400 mV (400m) on the popĆup display touch.............
Select Enter touch..........................................
Check that 400 mV has now entered into the display below Vert Offset:
M2.
ăStep 8:ăConnectthe VoltageSource to the DIRECT input of the CalibraĆ
tion Head.
ăStep 9:ă the VoltageSource to 400 mV.
ăStep 10:ăWaitfor averaging to complete. Then,  that DMean value
is between -2 mV and +2 mV.
ăStep 11:ăRepeat Steps 2 through 10 for the offset values shown in
Table 4Ć4.
ăStep 12:ăPowerĆoffthe CSA 803C.