Adjustments
Adjustment Procedures
5Ć20
This procedure verifies the correct operation and accuracy of the
CSA 803A's input/output systems(see Figure 5Ć6).
This procedure should be performed only if maintenance is required.
Its use is not necessary for verifying instrument specifications.
Measurement Limits
The measurement limit for the real time clock is 1,000,000 m5ms.

Setup to Examine/Adjust Real Time Clock

CSA 803C
CH1
10XProbe
A14
I/OBoard
(seeFigure 5Ć6)
ToTP310
ToGround
TestPoint
Frequency
Counter

Procedure to Examine/Adjust Real Time Clock

ăStep 1:ăInitialize the CSA 803C settings as follows:
UTILITYbutton press...........................................
Initialize (in major menu) touch..............................
Initialize (in verification popĆup menu) touch..................
Real Time Clock