6Ć38
%" %Remove the single Torxhead screws that secures the transĆ
former,on the A8 CRT Driver board, to the instrument chassis. This
screw is located at the rear of the transformer.The A8 CRT Driver board
is now not attached to the instrument chassis.
%" %Remove connectors J52, J56, and J57 from the A8 CRT Driver
board. Note the position of the multiĆpin connectors' index triangles to
ensure that you can correctly replace these connectors.
%" %Lift the A8 CRT Driver board partially out of the instrument
chassis.
%" %Remove connectors J54 and J55 from the A8 CRT Driver
board. Note the position of the multiĆpin connectors' index triangles to
ensure that you can correctly replace these connectors.
%" %Remove the A8 CRT Driver board and the A7 CRT Socket
board.
%" %Unsolder the two wires that connect the A8 CRT Driver board
to the A7 CRT Socket board (the two boards are now separated).
Toreplace the A8 CRT Driver board and the A7 CRT Socket board, perform
the previous steps in reverse order.
J53
J56
SolderPoints
#! &%$  "   " !