Performance Verification
CSA 803C Service Manual 4Ć3
Table4Ć2 (Test Equipment), contains suggested test equipment for use in
this manual. Procedure steps are based on the test equipment examples
given, but you may substitute other equipment with similar specifications.
The use of different equipment may alter test results, setup information, and
related connectors and adapters.
TableĂ4Ć2:ăTest Equipment
Description Minimum
Specification
Examples of Applicable
TestEquipment
High Frequency
Sine WaveGeneraĆ
tor
10MHzto10GHz WILTRON6759B Opt 2
Sampling Head SDĆSeries SDĆ24 and any other Tektronix
SDĆSeries Sampling Head
except SDĆ14
Digital Voltmeter
(w/test leads)
v0.005% Accuracy FLUKE 8842A Digital Voltmeter
FrequencyCounter One Part in
1,000,000
Accuracy
TektronixDC 5010 Universal
Counter/Timerwith a
TM 5000ĆSeries PowerModule
Calibration Head TektronixCalibration Head
067Ć1413Ć00
Precision Voltage
Source
DATAPRECISION 8200
TestTerminal any GPIB
(IEEEĆ1978)controlĆ
ler,or ASCII termiĆ
nal equipped with
an RSĆ232ĆC port.
RequirescompatĆ
ible RSĆ232ĆC serial
interfacecable
IBM compatible PC with termiĆ
nal emulation software
Coaxial Cable, 50 W
(3 required)
50 W,36Ćinch, male
BNCconnectors
TektronixPart 012-0482-00
Coaxial Cable, RF RF cable, 12 in.,
SMAconnectors
TektronixPart 174-1364-00
RF cable, 8.5 in.,
SMAconnectors
TektronixPart 174-1120-00
TestEquipment