Performance Tests
Performance Verification
4Ć46
ăStep 6:ăOn the CSA 803C, press the AUTOSETbutton.
ăStep 7:ăCheck the peakĆtoĆpeakamplitude as follows:
MEASUREbutton press........................................
Measurements(in the major menu) touch.....................
PeakĆPeak(in the popĆup menu) touch.......................
Rise (in the popĆup menu) touch.............................
Rise (in the major menu) touch..............................
Proximal (in the popĆup menu) touch.........................
Adjust Proximal for 20% and Distal for 80%.
 that peakĆtoĆpeakamplitude is within 212.5 mV and 287.5 mV
(250 mV ±15%).
Check that rise time is approximately 250Ăps.