Troubleshooting
MTG100 & MTG300 MPEG Generator Service Manual 6Ć65
Solid state disk
Connect a monitor to the VGA
connector on the rear panel or
short between pin 2 and 3 of J278
on the A40 MISC (Power
Distribution & Interface) board
(see Figure 6Ć36) to be able to
display the boot screen on the
LCD.
Set the boot drive to
A:then:C".
Power on the instrument and
display the BIOS setup screen.
Insert a DOS boot disk into the
floppy disk drive, and then
reboot the instrument.
After the instrument starts up in
the DOS environment, check
the c drive using the
Scandisk.exe command.
If any error is reported,
replace the solid state disk.
Connect a keyboard to the
KEYBOARD connector on the
rear panel.
Figure 6Ć35: Troubleshootingprocedure (12)