Performance Verification
4Ć10 MTG100 & MTG300 MPEG Generator Service Manual
17. In the Real Time Analyzer window, click the Stop Analysis toolbar button
to stop analysis.
18. Press the START/STOP button on the MPEG generator to stop the signal
output, and then remove the BNC cable from the SSI OUT connector.
This test confirms that transport stream data is output correctly in parallel format
from the ECL/TTL/LVDS OUT connector on the MPEG generator. The
following equipment and MPEG test signal is required for this test:
MPEG test system
Parallel interface cable
test64.trp MPEG test signal
Perform the following procedure to verify that transport stream data in parallel
format is output correctly from the ECL/TTL/LVDS OUT connector on the
MPEG generator:
1. Use the parallel interface cable to connect the ECL/TTL/LVDS OUT
connector on the MPEG generator to the LVDS/ECL/RS422 Level In
connector on the MPEG test system rear panel as shown in Figure 4-7.
2. Open the test64.trp file on the MPEG generator.
a. Select Open TS File... from the FILE pull-down menu.
b. In the resulting Select TS File dialog box, select the test64.trp file, and
then press the OK bezel button.
3. Select Level : ECL from the OUTPUT menu on the MPEG generator, and
then press the OK bezel button.
4. Select Interface : Parallel from the OUTPUT menu on the MPEG
generator, and then press the OK bezel button.
5. Press the START/STOP button on the MPEG generator to start the signal
output of the test64.trp file.
6. Verify that the following LEDs on the MPEG generator’s front panel are
illuminated: PLL, BUFFER, and HDD.
Parallel Output